Kodak X-Omat 480 RA Processor Installation Instructions Manual page 31

Hide thumbs Also See for X-Omat 480 RA Processor:
Table of Contents

Advertisement

[9]
Adjust the position of the BLOWER MOTOR
for correct alignment and tension of the BELT.
The BELT has the correct tension if no
loud noises occur from the BELT when
the PROCESSOR is energized.
[10]
Tighten the 3 BOLTS.
Do not place hands or fingers near the moving parts.
[11]
Move the wall power switch to the ''ON''
position.
[12]
Energize the PROCESSOR by moving the
MAIN CIRCUIT BREAKER CB1 and the
AUXILIARY CIRCUIT BREAKERS CB2, CB3,
and CB4 to the ''I'' position.
[13]
Check for correct tension and alignment of the
BELT.
[14]
If it is necessary to adjust the BLOWER
MOTOR:
• Deenergize the PROCESSOR by moving
the MAIN CIRCUIT BREAKER CB1 to the
''O'' position.
• Make the adjustment.
[15]
Do steps ?? through ?? until the BLOWER
MOTOR is adjusted correctly.
[16]
To continue with the remaining steps,
deenergize the PROCESSOR by moving the
MAIN CIRCUIT BREAKER CB1 to the ''O''
position.
[17]
Install the BLOWER MOTOR COVER.
[18]
Install the FEED-END, MIDDLE ACCESS
PANEL onto the PROCESSOR.
[19]
At the back of the REPLENISHMENT PUMPS
B3 and B4, pull out and move the SWITCHES
to the 50 Hz position.
It might be necessary to loosen the
hardware and move the
REPLENISHMENT PUMPS to access
the SWITCHES.
1C0932
NOTE
NOTE
INCORRECT
Figure 36 Aligning the Blower Motor Pulley
REPLENISHMENT
PUMP B3
Figure 37 Setting the Switch to 50 Hz
CORRECT
H108_0074CA
REPLENISHMENT
PUMP B4
SWITCHES
H108_0292ACA
H108_0292AA
31