[5]
Install the TOP COVER of the PROCESSOR.
[6]
Check the PROCESSOR for leakage in the following areas:
(a)
RECIRCULATION PUMP FITTINGS
(b)
REPLENISHMENT PUMP CHECK VALVES
(c)
TANK FITTINGS
(d)
THERMOWELL ASSEMBLIES
(e)
PLUMBING FIXTURES
(f)
DRAIN VALVES
[7]
Check that the OVERFLOW HOSES are draining water from the TANKS.
[8]
Check that the DEVELOPER REPLENISHMENT HOSE and the FIXER REPLENISHMENT HOSE are
connected to the correct plumbing fittings.
[9]
Check that the correct REPLENISHMENT PUMP is operating.
Miscellaneous Components Check
[1]
Check that the following components operate:
(a)
DRYER BLOWER MOTOR
(b)
WATER SOLENOID - Listen for the sound of water entering the PROCESSOR.
(c)

EXHAUST FAN.

Display Panel Check
[1]
Check that the green ''Ready'' light on the DISPLAY PANEL illuminates after the PROCESSOR has
been allowed to warm-up for approximately 15 minutes at 26°C (72°F) ambient conditions and 26°C
(72°F) water temperature.
[2]
Check that the DEVELOPER temperature illuminated on the DISPLAY PANEL is in the operating range
between 35° and 38.3°C (95° and 101°F).
Transport Check
Feed one sheet of processed 35 x 43 cm (14 x 17 in.) film into the PROCESSOR to check that the transport
system and the REPLENISHMENT PUMPS operate correctly.