4.2.1 Prepare the patient

A click on the Prepare button will display the calibration screen and turn on the calibration LED's.
Figure 4-3 Calibration Preparation Screen
4.2.1.1

Adjust the eye/camera

Instruct the patient to fixate the center luminous dot.
Align the camera by rotating it in its ball-and-socket joint such that the eye is finally centered in the
image. The camera can be rotated in three directions: yaw, pitch, and roll. At one time, however,
rotate the camera only in one of these three possible directions. After the eye is centered in one
direction, e.g., in the horizontal direction, choose the next rotation direction, e.g., the vertical direction.
A possible alignment sequence to center the eye is shown Figure 4-4.
Figure 4-4 Centering the pupil in the image with a sequence of three camera rotations. Immediately after the goggles are placed on
the patient, the eye can assume a position as in the upper left image. From this starting position, the camera first needs to be
rotated such that the pupil is centered horizontally. The red arrow in the top left image indicates how the pupil will move during such