Returns
<NR1>

*PSC

This order control if power supply send a query or not when it is reset.
1 OR ON: When power supply is reset, status byte register operation event
enable register, query event enable register and standard event status register
are all reset.
0 OR OFF: The data of status byte register, operation event enable register,
quest event enable register and standard event status enable register is stored
in nonvolatile register, and is recalled when power supply is reset.
Command syntax
*PSC <bool>
Arguments
0|1|ON|OFF
Query syntax
*PSC?
Returns
0|1
Related Command
*ESE *SRE STAT:OPER:ENAB STAT:QUES:ENAB

*RST

This command reset the power supply to default settings.
CAL:SEC:STAT 1
OUTP OFF
CURR MAX
VOLT:LIMIT MAX
INST:COM:OFF
Command syntax
*RST>
Arguments
None

*SRE

This command can set the parameter of state byte enable register.setting
parameter can determine which byte value of state byte register is 1 and the
Copyright© Itech Electronics Co., Ltd.