CALibrate:SECure[:STATe]

This command is used to set the calibration state.
Command syntax
CALibration:SECure:[STATe] {<ON|OFF>,[<password>]}
Arguments
0|1, '6302'
Example
CAL:SEC 0, '6302'; CAL:SEC 1
Query syntax
CALibration:SECure:STATe?
Arguments
None

CALibrate:VOLTage:LEVel

This command is used to
Command syntax
CALibrate:VOLTage:LEVel <level>
Arguments
P1|P2|P3|P4
Note: Parameters must be sent by sequence of P1->P2->P3->P4.

CALibrate:VOLTage[:DATA]

Return the actual output voltage value at present calibration point of power
supply.
Command syntax
CALibrate:VOLTage[:DATA] <NRf>
Arguments
<NRf>
Example
CAL:VOLT 0.3V
Note: Please execute this order when power supply output is stable.

Chapter1 Calibration Commands

define the voltage calibration point.
Copyright© Itech Electronics Co., Ltd.