Status:questionable:instrument:enable; Status:questionable:instrument:isummary1[:Event]; Status:questionable:instrument:isummary1:Enable; Status:questionable:instrument:isummary1:Condition - ITech CalPower IT6302 Programming Manual

Triple output programmable dc power supply
Hide thumbs Also See for CalPower IT6302:
Table of Contents

Advertisement

Returns
<Nrf>

STATus:QUEStionable:INSTrument:ENABle

This command can re-write the parameter of query enable register in the
channel.
Command syntax
STATus:QUEStionable:INSTrument:ENABle
Arguments
Register parameter
Query syntax
STATus:QUEStionable:INSTument:ENABle?
STATus:QUEStionable:INSTrument:ISUMmary1[:EVEN
t]?
This command can query the event register in channel 1. When the query value
is returned, the register is reset.
Query syntax
STATus:QUEStionable:INSTrument:ISUMmary1 [:EVENt]?
Arguments
None
STATus:QUEStionable:INSTrument:ISUMmary1:ENABl
e
This command can re-write the parameter of query enable register in channel
1.
Command syntax

STATus:QUEStionable:INSTrument:ISUMmary1:ENABle

Arguments
Register parameter
Query syntax
STATus:QUEStionable:INSTument:ISUMmary1:ENABle?
STATus:QUEStionable:INSTrument:ISUMmary1:COND
ition?
This command can query the parameter of query status register in channel 1.