Performing Temperature Calibration - Emerson Micro-Motion 2400S Configuration And Use Manual

Transmitters with analog outputs
Table of Contents

Advertisement

Measurement Performance
10.6

Performing temperature calibration

Temperature calibration is a two-part procedure: temperature offset calibration and temperature slope
calibration. The entire procedure must be completed without interruption.
You can calibrate for temperature with ProLink II software. See Figure 10-16.
Figure 10-16
Temperature calibration – ProLink II
Temperature Offset calibration
ProLink Menu >
Configuration and Use Manual
Fill sensor with low-
temperature fluid
Wait until sensor achieves
thermal equilibrium
Calibration >
Temp offset cal
Enter temperature of low-
temperature fluid
Do Cal
Calibration in Progress
light turns red
Calibration in Progress
light turns green
Close
Temperature Slope calibration
Fill sensor with high-
temperature fluid
Wait until sensor achieves
thermal equilibrium
ProLink Menu >
Calibration >
Temp slope cal
Enter temperature of high-
temperature fluid
Do Cal
Calibration in Progress
light turns red
Calibration in Progress
light turns green
Close
Done
129