Emerson Micro-Motion 2400S Configuration And Use Manual page 35

Transmitters with analog outputs
Table of Contents

Advertisement

Flowmeter Startup
Figure 5-3
ProLink II – Loop test procedure
Fix Milliamp 1/2
Enter mA value
Fix mA
Read output at
receiving device
Correct?
Yes
Unfix
Figure 5-4
Communicator – Loop test procedure
1 Fix Analog Out 1/2
4 mA
20 mA
Other
Configuration and Use Manual
ProLink Menu >
Test
Fix Freq Out
Fix Discrete Output
Enter Hz value
Fix Frequency
Read output at
receiving device
receiving device
Correct?
No
Check output wiring
Troubleshoot receiving device
On-Line Menu >
3 Diag/Service >
2 Loop test
2 Fix Frequency Out
10 kHz
Other
Read output at
Read output at
receiving device
receiving device
Correct?
Correct?
Yes
Loop test successful
Check output wiring
END to unfix
Troubleshoot receiving device
Read Discrete Input
Toggle remote input
ON or OFF
Verify state at
Verify state
at transmitter
Correct?
Correct?
Yes
Loop test successful
3 Fix Discrete Out 1
ON or OFF
Verify state at
receiving device
Correct?
No
device
No
Check input wiring
Troubleshoot input device
27