Emerson Micro-Motion 2400S Configuration And Use Manual page 135

Transmitters with analog outputs
Table of Contents

Advertisement

Measurement Performance
Figure 10-14
D3 or D3 and D4 density calibration – ProLink II
Close shutoff valve
downstream from sensor
Configuration and Use Manual
D3 calibration
Fill sensor with D3 fluid
ProLink Menu >
Calibration >
Density cal – Point 3
Enter density of D3 fluid
Do Cal
Calibration in Progress
light turns red
Calibration in Progress
light turns green
Close
D4 calibration
Fill sensor with D4 fluid
ProLink Menu >
Calibration >
Density cal – Point 4
Enter density of D4 fluid
Do Cal
Calibration in Progress
light turns red
Calibration in Progress
light turns green
Close
Done
Done
127