Section 4 Operation
Stopping
the Unit

Power Down

Shut Down

CAUTION
CAUTION
CAUTION

Restarting

4-24
• Press
to the left of the stop symbol,
and the stop symbol will turn into a start symbol,
Menu
• Press
. The circulator's display will go blank and the blue LED
will illuminate indicating the unit is in the standby mode.
• Place the circuit protector on the rear of the immersion circula-
tor to the O position. The blue LED will extinguish.
• On refrigerated units, place the circuit protector on the rear of
the bath to the O position.
Using any other means to shut the unit down can reduce the life of
the compressor.
Always turn the unit off and disconnect it from its supply voltage
before moving the unit.
The circuit protector(s) located on the rear of the component(s) is
not intended to act as a disconnecting means.
NOTE When quickly restarting refrigerated units, the compressor may
take up to 10 minutes before it starts to operate.
20.00°C
. The unit will stop
.
SP1
20.00C
I
O