Thermo Scientific Premium PC 200 Manual page 56

Immersion/ refrigerated bath/ heating bath circulators
Hide thumbs Also See for Premium PC 200:
Table of Contents

Advertisement

Section 6 Troubleshooting
6-2
Low Fixed Temp.
Low Temperature
Motor Fault
MOL
Motor Overload
Open RTD1
Internal
Open RTD2
External
Shorted RTD1
Internal
Shorted RTD2
External
Warning Displays
The unit will continue to run with a warning. A warning also sounds the
alarm, if enabled.
High Temperature
High Temperature
Low Level
Low Temperature
unit's nonadjustable low temperature protection
limit exceeded
check fluid selection
adjustable high temperature protection limit
exceeded
check limit setting
check fluid selection
it can take over 10 minutes for the motor
temperature to get low enough before the unit
can be restarted
allow unit to cool down
open internal temperature sensor
open external temperature sensor
shorted internal temperature sensor
shorted external temperature sensor
adjustable high temperature protection limit
exceeded
check limit setting
check fluid selection
adjustable high temperature protection limit
exceeded
check limit setting
check fluid selection
reservoir fluid level too low for safe operation
check fluid level
adjustable low temperature protection limit
exceeded
check limit setting
check fluid selection
Thermo Scientific