Advanced - Ramping
Advanced
Thermo Scientific
Ramping
is used to build/edit a ramp program.
Ramping
Ramp
1
Steps
2
Cycles
1
Variance
0.05°C
Ramps
There can be 10
Cycled
up to 100 times.
Cycles
sets the number of times the entire ramp program will be re-
peated after the last step is completed. For example, selecting 3
runs the entire ramp program a total of 3 times.
Variance
The
is used to set a temperature range, the program will
start when the fluid temperature is within this range. For example, if the
Start Temp
desired
program will automatically start when the bath temperature is between
20°C to 30°C.
The right side of the display is used to build each step. Use the
key to sequence through the steps.
Start Temp
NOTE The
step 1, and so on.
Ramping
Steps
, each with 30
Variance
is 25°C and the
for a step 2 will be the
Section 4 Operation
Step
1
Start Temp
40.00°C
End Temp
20.00°C
Duration
1min
Ramp
. Each
can be
Cycles
is set to +5°C, the
End Temp
from
4-19