Installation; Ambient Conditions - Thermo Scientific Premium PC 200 Manual

Immersion/ refrigerated bath/ heating bath circulators
Hide thumbs Also See for Premium PC 200:
Table of Contents

Advertisement

Ambient
Conditions
CAUTION
CAUTION
CAUTION
Thermo Scientific

Installation

Section 3
Ambient Temperature Range
Maximum Relative Humidity
Operating Altitude
Overvoltage Category
Pollution Degree
Degree of Protection
The unit is designed for continuous operation and for indoor use.
The units normally ships with the immersion circulator mounted facing the
reservoir. You may change the position ±90° by removing thumb screws, no
tools are required.
Do not mount it backwards; the line cord could contact the reservoir
fluid.
Rotate immersion
circulator only ±90°
from shown position
Never place the unit in a location where excessive heat, moisture,
inadequate ventilation, or corrosive materials are present.
Refrigerated units should be left in an upright position for 24 hours at
room temperature before starting. This will ensure the lubrication oil
has drained back into the compressor.
5°C to 40°C (41°F to 104°F)
80% at 31°C (88°F)
Sea Level to 2000 meters (6560 feet)
II
2
IP 20
3-1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents