Fluids Type
is used to identify the type of fluid used. The controller
uses the fluid type to automatically set certain operating parameters.
Fluid Type
SYNTH
SIL
Press
or
NOTE For high viscosity fluids with a wide "Heating-up range" (e.g.,
SIL300 - see Section 3), the controller automatically slows the heating
from ambient to the "Operating temperature range" to help prevent fluid
degradation that is caused by rapid heating while still at high viscosities.
Fluid Limits °C:
Min,
SYNTH 60
-50
SYNTH 200
30
SYNTH 260
45
SIL 100
-75
SIL 180
-40
SIL 200
20
SIL 300
80
Water
5
EG/Water
-30
Other
-90
Fluid Type
next to the desired fluid or fluid type.
Max
45
210
250
75
200
200
300
95
100
300
Water
Glycol-Water
Other