Appendix
Table 2
QUALIFIER BYTE
00
0 precision (1), no units of measure
10
1 precision (0.1), no units of measure
20
2 precision (0.01), no units of measure
11
1 precision (0.1), °C
21
2 precision (0.01), °C
12
1 precision (0.1), °F
22
2 precision (0.01), °F
1D
1 precision (0.1), °K
2D
2 precision (0.01), °K
Example: The temperature of 45.6°C would be represented by the qualifier 11 hex, followed by the 2 bytes 01
C8 hex (456 decimal).
Table 3 REQUEST STATUS (5 bytes)
BIT
d1
b.7
RTD1 Open
Error
b.6
RTD1 Short
Error
b.5
RTD1 Open
Fault
b.4
RTD1 Short
Fault
b.3
0
b.2
0
b.1
0
b.0
0
Unit of Measure (not all used)
Index
Unit
0
NONE
1
Temperature in °C
2
Temperature in °F
3
Flow liters per minute
4
Flow in gallons per minute
5
Time in seconds
6
Pressure in PSI
7
Pressure in bars
6
d2
d3
RTD2 Open
High temp
Error
fixed fault
RTD2 Short
Low temp
Error
fixed fault
RTD2 Open
High temp
Warn
fault
0
Low temp
fault
RTD2 Open*
Low level fault
RDT2 Short*
High temp warn
High RA temp
Low temp warn
fault
HTC fault
Low level warn
*Ignore this display
Unit of Measure (not all used)
Index
Unit
8
Resistivity in MΩ-cm
9
%
10
Volts
11
Pressure in kPa
12
Conductivity in µS/cm
13
Temperature in °K
d4
d5
Beeper on
External Sensor
controlling
0
Heat icon
flashing
Fault
Heat icon on
Unit stopping
Cool icon
flashing
Unit on
Cool icon on
Pump on
HPC fault
Compressor on
MOL fault
Heat output > 0
Pump speed fault
Thermo Scientific