Error Displays

Thermo Scientific

Troubleshooting

Section 6
Error messages are cleared by pressing the enter key. Once the cause of the
error message is identified and corrected, to restart the unit - on units with
an HTC error press the reset key full down - and then for all units press the
enter key again. If the message does not clear contact our Sales, Service and
Customer Support.
Fault Displays
The heating element, pump and, if applicable, refrigeration shut down with a
fault. A fault also sounds the alarm, if enabled.
Message
High Fixed Temp.
High Temperature
High Temperature
Refrigeration
HPC
High Press. Cutout
HTC
High Temp. Cutout
LLC
Low Level Cutout
FAULT:
HTC
HIGH TEMP. CUTOUT-
PRESS ENTER
Error Message (Typical)
Action
unit's nonadjustable high temperature protection
limit exceeded
check fluid selection
check environmental conditions
adjustable high temperature protection limit
exceeded
check limit setting
check fluid selection
ensure unit has adequate ventilation
check voltage supply
the refrigeration may need servicing
check for obstructions to air flow
the refrigeration may need servicing
high temperature protection limit exceeded
check limit setting
allow unit to cool down
when operating at high temperatures ensure
pump is in high-speed mode
reservoir fluid level too low for safe operation
check fluid level
check for leaks
6-1