Measured/Computed Data - YOKOGAWA Vigilantplant mR10000 User Manual

Communication interface
Table of Contents

Advertisement

5.2 Output Format of ASCII Data

Measured/computed Data

5-8
• The FD (FD0) or FY (FYInst, FYTlog1, or FYTlog2) command is used to output the
data.
• Syntax
The measured/computed data are output in the following syntax along with the date
and time information for each channel
EACRLF
DATE_yy/mo/ddCRLF
TIME_hh:mi:ss.mmmt_S
s_kcca
a
a
a
uuuuuuf
1
2
3
4
••••••••••••••••••••••••••••
ENCRLF
yy
mo
dd
hh
mi
ss
mmm
t
S
S
S
S
S
S
1
2
3
4
5
6
S
1
S
2
S
3
S
S
S
4
5
6
s
k
cc
a
a
a
a
1
2
3
4
S
S
S
S
S
CRLF
1
2
3
4
5
6
dddddEf
ppCRLF
1
2
Year (00 to 99)
Month (01 to 12)
Day (01 to 31)
Hour (00 to 23)
Minute (00 to 59)
Second (00 to 59)
Millisecond (000 to 999. A period is placed between seconds
and milliseconds.)
'S'(=summer) or ' ' space(=winter)
Data status
Takes on the values below for the FYTlog1 and FYTlog2
commands. For all other cases, they are all spaces.
Time change during TLOG computation: T (occurred) or space
(not occurred)
Power OFF and ON during TLOG computation: P (occurred) or
space (not occurred)
Data reset during the TLOG computation: R (occurred) or
space (not occurred)
All spaces
Channel data status (N, D, S, O, B, E)
N: Normal
D: Differential input
S: Skip
O: Over
B: Burnout
E: Error
Channel type
0: Measurement channel
A: Computation channel
Channel number
µR10000: 01 to 06, 0A to 0P
µR20000: 01 to 24, 0A to 1P
a
Alarm status (level 1)
1
Alarm status (level 2)
a
2
a
Alarm status (level 3)
3
a
Alarm status (level 4)
4