Command Syntax
ESC O_xx CRLF
(_ space)
ESC C_xx CRLF
(_ space)
* The causes that the condition become "The instrument with the specified address does not
exist" is such as a command error, the address not matching that of the instrument, the
instrument is not turned ON, and the instrument not being connected via the serial interface.
• The "xx" in the table indicates the instrument's address. Specify the address that is
assigned to the instrument from 01 to 32.
• Only one instrument can be opened at any one time.
• When an instrument is opened with the ESC O command, all commands on the
instrument become active.
• When an instrument is opened with the ESC O command, any other instrument
that is open is automatically closed.
• Use CR+LF for the terminator. LF cannot be used.
Note
The ASCII code of ESC is 1BH.
Meaning
Response
Open the instrument.
• Response from the instrument with the
specified address
ESC O xx CRLF
• Response when the instrument with the
specified address does not exist*
Close the instrument.
• Response from the instrument with the
specified address
ESC C xx CRLF
• Response when the instrument with the
specified address does not exist*