Decimal Point Position/Unit Information - YOKOGAWA Vigilantplant mR10000 User Manual

Communication interface
Table of Contents

Advertisement

Decimal Point Position/Unit Information

IM 04P01B01-17E
• The FE command (FE1) is used to output the data.
• Syntax
The data is output for each channel in the following syntax.
EACRLF
s_kccuuuuuu,ppCRLF
••••••••••••••••••
ENCRLF
Data status (N, D, S)
s
N: Normal
D: Differential input
S: Skip (When the input range is set to SKIP for a measurement
channel or when the channel is turned OFF for a computation
channel)
k
Channel type
0: Measurement channel
A: Computation channel
Channel number
cc
µR10000: 01 to 06, 0A to 0P
µR20000: 01 to 24, 0A to 1P
uuuuuu
Unit information (6 characters, left-justified)
mV____: mV
V_____: V
^C____: °C
xxxxxx: (User-defined character string)
Decimal point position (00 to 04)
pp
No decimal (00000) for 00.
One digit to the right of the decimal (0000.0) for 01.
Two digits to the right of the decimal (000.00) for 02.
Three digits to the right of the decimal (00.000) for 03.
Four digits to the right of the decimal (0.0000) for 04.
_
Space
• Example
EA
N 001mV
,01
N 002mV
,01
EN
5.2 Output Format of ASCII Data
5
5-7

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents