Control Commands - YOKOGAWA Vigilantplant mR10000 User Manual

Communication interface
Table of Contents

Advertisement

4.6 Control Commands

4.6
Control Commands
DS
Switches the execution mode.
Syntax
DS p1<terminator>
p1 Mode
0 Run mode
1 Basic Setting mode
Example
Switch to Basic Setting mode.
DS 1
Description The execution mode cannot be changed to Basic
Setting Mode while recording or computation is
in progress.
PS
Starts/Stops recording.
Syntax
PS p1<terminator>
p1 Start/Stop recording.
0 Start
1 Stop
Example
Start recording.
PS 0
UD
Switches the screen/switches
the channel.
Returning to the data display screen
Returns from the Setting mode screen, the screen that
appears when the FUNC key is pressed, or the screen that
appears when the DISP MENU key is pressed back to the
measurement data display screen.
Syntax
UD p1<terminator>
p1 Fixed to 0.
Example
Return to the data display screen.
UD 0
Displaying the specified data display screen
Syntax
UD p1,p2<terminator>
p1 Fixed to 1.
p2 Screen number (1 to 15)
Example
Display data display screen 2.
UD 1,2
Switching the display channel (manual switching)
Syntax
UD p1<terminator>
p1 Fixed to 2.
Example
Switch the display channel.
UD 2
Description • An error results if the specified screen is set to
SKIP for UD1.
• UD2 is valid on screens whose display
channel is set to manual switching.
AK
Executes alarm acknowledge
(Alarm ACK).
Syntax
AK p1<terminator>
p1 Executes alarm acknowledge (0)
Example
Execute alarm acknowledge.
AK 0
4-28
TL
Starts/stops/resets computation
(/M1 option).
Syntax
TL p1<terminator>
p1 Operation type
0 Computation start
1 Computation stop
2 Computation reset
Example
Start the computation.
TL 0
MP
Starts/Stops manual print.
Syntax
MP p1<terminator>
p1 Operation type
0 Manual print start
1 Manual print stop
Example
Start manual print.
MP 0
LS
Starts/Stops the list (setting
information) printout.
Syntax
LS p1<terminator>
p1 Operation type
0 List print start
1 List print stop
Example
Start list print.
LS 0
Description List print refers to printing of settings of Setting
mode.
SU
Starts/Stops the setup list (basic
setting information) printout.
Syntax
SU p1<terminator>
p1 Operation type
0 Setup list print start
1 Setup list print stop
Example
Start setup list print.
SU 0
Description Setup list print refers to printing of settings of
Basic Setting mode.
MS
Prints the message.
Syntax
MS p1<terminator>
p1 Message number (1 to 5)
Example
Print the message of message number 3.
MS 3
Description The message string is set with the SG command.
AC
Clears the alarm printout buffer.
Syntax
AC p1<terminator>
p1 Clear the alarm printout buffer (0)
Example
Clear the alarm printout buffer.
AC 0
IM 04P01B01-17E

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents