Output Commands - YOKOGAWA Vigilantplant mR10000 User Manual

Communication interface
Table of Contents

Advertisement

MC
Clears the message printout
buffer.
Syntax
MC p1<terminator>
p1 Clear the message printout buffer (0)
Example
Clear the message printout buffer.
MC 0
VG
Resets the report data of the
periodic printout.
Syntax
VG p1<terminator>
p1 Operation type
2 Reset the report data of the periodic
printout.
Example
Reset the report data of the periodic printout.
VG 2
Description This setting is valid when the recorder is set to
print the report data using periodic printout.
YC
Initializes settings.
Syntax
YC p1<terminator>
p1 Initialization type
0
Initialize the Setting mode and Basic
Setting mode settings.
Initialize the Setting mode settings.
1
Example
Initialize the Setting mode and Basic Setting
mode settings.
YC 0
UY
Stops the record position
adjustment.
Syntax
UY p1<terminator>
p1 Fixed to 0.
Query
UY?
Example
Stop the record position adjustment.
UY 0
Description Returns the execution status of the record
position adjustment in response to a query.
0: Stopped, 1: In progress
IM 04P01B01-17E
4.6 Control Commands/4.7 Output Commands
4.7

Output Commands

BO
Sets the byte output order.
Syntax
BO p1<terminator>
p1 Byte order
0 Outputs the data MSB first.
1 Outputs the data LSB first.
Query
BO?
Example
Output the data MSB first.
BO 0
Description This command applies to the byte order of
numerical data during BINARY output.
CS
Sets the checksum (/C3 option).
Syntax
CS p1<terminator>
p1 Use/Not use the checksum
0 Not use
1 Use
Query
CS?
Example
Use the checksum.
CS 1
Description This command can be used only on the RS-
422A/485 communication interface.
IF
Sets the status filter.
Syntax
IF p1<terminator>
p1 Status filter value
(0.0.0.0 to 255.255.255.255)
Query
IF?
Example
Set the status filter value to 1.0.4.0.
IF 1.0.4.0
Description For details, see chapter 6.
Note
Initialization of BO/CS/IF Command Settings
For serial communications
Settings entered using the BO/CS/IF commands
revert to their initial values when the recorder is reset
(when the recorder is power cycled, or the user exits
the basic setting mode with the YE command).
Byte output order, checksum: 0
Status filter: 255.255.255.255
If the recorder is reset, you must restore these
settings.
For Ethernet communications
Settings entered using the BO/IF commands revert to
their initial values when the connection to the recorder
is cut. After reconnecting the recorder, you must
reenter the settings.