Infrared Link Communication Specifications; Communication Timeout; Communication Environment; Other Items To Be Installed - Canon AX-C4343W Technical Document

Digital x-ray imager 2.4 ghz/5 ghz
Table of Contents

Advertisement

Appendix — Technical description —

5.2 Infrared LINK communication specifications

This link action establishes infrared link communication between the infrared
communication section of the detector and the ready indicator connected to the image-
capture computer.
5.2.1
3 sec (The period during which the detector tries infrared communication after the
detector's POWER switch is pushed down.)
5.2.2
Communication range: Within 30 cm
Environment conditions: Cannot be used in an environment in which infrared radiation is

6 Other items to be installed

Use items selected by your sales representative for the following units:
• Wireless LAN access point
• LAN cable
• Micro USB cable (used to connect the ready indicator)
• Image-capture computer

7 Installation, repair and maintenance check

7.1 License for service engineer

Service technicians who have undergone training for servicing this product, including
specifications, installation procedures and parts replacement procedures. Limited to
those with permission from the department that hosts training.
Please never perform the installation, repair, maintenance check except an above service
engineer, for the product works safety and enough performance.

7.2 Installation procedures

Work-related precautions
Bear the following points in mind when installing the system.
(1)
Carry out the work while ensuring that none of the portable units will be dropped or
subjected to impact.
(2)
Ensure that the installation location satisfies the following criteria:
1) A location where the product will not be splashed with water
2) A location where the air does not contain moisture, dust, salt, sulfur or other such
substances and where the product will not be adversely affected by poor ventilation,
the air pressure level, the temperature and other such factors
3) A location which is not tilted or subject to vibration or impact (this includes the
locations when the product is conveyed)
4) A location where chemical products are not stored and where no gases are given off
82

Communication timeout

Communication environment

blocked.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents