Warnings And Important Information; Warnings - GE Invenia ABUS 2.0 Technical Publication

Table of Contents

Advertisement

Replacement Procedures

Warnings and important information

Warnings

WARNING
WARNING
WARNING
WARNING
8-2
Energy Control and Power Lockout for Invenia ABUS 2.0.
When servicing parts of the Invenia ABUS 2.0 where there is
exposure to voltage greater than 30 volts:
1. Follow LOCKOUT/TAGOUT procedures.
2. Turn off the Power Mains Switch.
3. Unplug the Invenia ABUS 2.0.
4. Maintain control of the Invenia ABUS 2.0 power plug.
5. Wait for at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
6. Remove/disconnect the battery, if present.
Invenia ABUS 2.0 components may be energized.
Because of the limited access to cabinets and equipment in the
field, placing people in awkward positions, GE has limited the
lifting weight for one person in the field to 35 LBS (16 KG).
Anything over 35 LBS (16 KG) requires 2 people.
At least two people are needed when replacing casters
(wheels) or adjusting brakes.
Avoid contacting used parts before disinfecting.