Cineloop Check (QG)

For activating Cine, creating and storing Cine Loops, and information on the Cine Timeline, refer to the
LOGIQ™ S6 Quick Guide.
4-3-7

Backend Processor Checks

If all the previous tests have been passed successfully, the backend processor is most likely OK.
If the system seems to be operating erratically, please refer to Chapter 7, Diagnostic/
Troubleshooting.
4-3-8

Probe/Connectors Usage (QG)

The LOGIQ™ S6 Quick Guide, provides information on connecting, activating, deactivating and
disconnecting probes.
4-3-9

Peripheral Checks

Check that peripherals work as described below:
Table 4-40 Peripheral checks
Step
1.
2.
4-3-9-1
Peripheral/Option Checks
If any peripherals or options are not part of the system configuration, the check can be omitted. Refer
to the User Manual for a list of approved peripherals/options.
Table 4-41 GE Approved Peripheral/Hardware Option Functional Checks
Step
1
2
B/W Printer Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
3
Color Printer Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
4
DICOM
5
InSite/iLinq
6
Camera
7
Footswitch
8
9
3D Probe
3
Task to do
Press Freeze
Press (P1) or (P4) on the Control panel
Item
VCR
Verify record/playback capabilities of the VCR. Clean heads and covers if necessary.
Verify that DICOM is functioning properly. Send an image to a DICOM device.
Verify that InSite is functioning properly. Ensure two-way remote communications.
(Warranty & Contract Customers only)
Verify hardcopy output of the film camera. Clean as necessary.
Verify that the footswitch is functioning as programed. Clean as necessary.
ECG
Verify basic operation with customer
Ask Engineering about Calibration Check for 3-D
Section 4-2 - General Procedure
LOGIQ™ S6 S
Expected Result(s)
Stop image acquisition.
The image displayed on the screen is
printed on B&W or Color printer depending
on the key assignment configuration.