Required facility needs (continued)
Desirable features
Ultrasound System
Direction 5444964-100 English Rev. 5
•
Dedicated single branch power outlet of adequate
amperage, meeting all local and national codes, which is
located less than 2.5 m (8 ft.) from the Ultrasound system's
proposed location
•
Door opening is at least 76 cm (30 in) wide
•
Proposed location for Ultrasound system is at least 0.5 m
(1.5 ft.) from the wall for cooling
•
Power outlet and place for any external peripheral are within
2 m (6.5 ft.) of each other with peripheral within 1 m of the
Ultrasound system to connect cables.
•
Power outlets for other medical equipment
•
Power outlets for test equipment within 1 m (3.2 ft.) of
Ultrasound system
•
Clean and protected space to store probes (in their cases or
on a rack)
•
Material to safely clean probes (done with a plastic
container, never metal)
For the amperage requirements, see: 'Electrical requirements'
on page 3-5.
•
Door is at least 92 cm (3 ft.) wide
•
Circuit breaker for dedicated power outlet is easily
accessible
•
Sink with hot and cold water
•
Receptacle for bio–hazardous waste, like used probe
sheaths
•
Emergency oxygen supply
•
Storage for linens and equipment
•
Nearby waiting room, lavatory, and dressing room
•
Dual level lighting (bright and dim)
•
Lockable cabinet ordered by GE for its software and
proprietary manuals
–
Common Service Information
Facility needs
3-25