GE LOGIC C2 Series Technical Publication page 88

Ultrasound system
Table of Contents

Advertisement

Care and Maintenance
Step
Item
11.
Wheels and
Brakes
12.
Cables and
Connectors
13.
Shielding and
Covers
14.
Control Panel
15.
Probe Holders
16.
Power and
System Status
Indicators
17.
Battery
18.
External
Microphone
4-10
Table 4-5:
Physical checks (Continued)
• Where applicable, check all wheels and casters for wear and verify
operation of foot brake, to stop the Ultrasound system from moving, and
release mechanism.
• Where applicable, check all wheel locks and wheel swivel locks for
proper operation.
Check all internal cable harnesses and connectors for wear and secure
connector seating. Pay special attention to footswitch assembly and
probe strain or bend reliefs.
Check to ensure that all EMI shielding, internal covers, air flow panels and
screws are in place. Missing covers and hardware could cause EMI/RFI
problems while scanning.
Inspect alphanumeric keyboard and Operating Panel. Record any
damaged or missing items.
Where applicable, inspect the Probe Holders for cracks or damage.
Check for proper operation of all Power and System Status Indicators.
Where applicable, check that the battery is not damaged, does not leak,
does not emit an odor, and is not deformed or discolored. Observe all
warnings and cautions for battery handling, recharging, storing, and/or
disposal,
Where applicable, check for proper operation of any external microphones
by recording an audio test.
Ultrasound System
Description
Common Service Information
Direction 5444964-100 English Rev. 5