Data sheet for enclosure/chassis leakage current
NOTE:
Table 4-10:
Unit under test____________________________________
Test Conditions
System
Grounding/
Power
off
closed
off
open
on
closed
on
open
NOTE:
Table 4-11:
Unit under test____________________________________
Test Conditions
System
Grounding/
Power
PE
off
closed
off
open
on
closed
on
open
off
closed
(reversed
polarity)
Ultrasound System
Direction 5444964-100 English Rev. 5
Table 4-10 shows a typical format for recording the enclosure/
chassis leakage current. Measurements should be recorded
from multiple locations for each set of test conditions. The actual
location of the test probe may vary by Ultrasound system.
Record all data in the Electrical safety tests log.
Not all test procedures are applicable to all areas of the world.
Reversed Polarity testing content satisfies regions following IEC
62353:2007 and IEC 60601-1:2005.
Typical data format for recording enclosure/chassis leakage
Potential
equilibrium
PE
connectorl
Values in italics font are given as examples only.
Typical data format for recording enclosure/chassis leakage
Potential
Limit
equilibrium
µA
connector
100
500
100
500
100
Common Service Information
Date of test:_____________
Measurement/Test Point Location
Lower
Frame
Connector
Date of test:_____________
Measurement/Test Point Location
Monitor
housing
Electrical safety tests
Probe
Main
Handle
Probe
connector
4-27