Site Preparations
Environmental specifications for Ultrasound System system
Cooling
NOTE:
Lighting
3-4
Refer to the User Manual/ Basic User Manual for the product.
The cooling requirement for a console Ultrasound system with
monitor and on board peripherals, is up to 3800 BTU/h. This
figure does not include cooling needed for lights, people, or
other equipment in the room.
Each person in the room places an additional 300 BTU/h
demand on the cooling system.
Bright light is needed for Ultrasound system installation, updates
and repairs. However, operator and patient comfort may be
optimized if the room light is subdued and indirect. Therefore a
combination lighting system (dim/bright) is recommended. Keep
in mind that lighting controls and dimmers can be a source of
EMI which could degrade image quality. These controls should
be selected to minimize possible interference.
Ultrasound System
–
Common Service Information
Direction 5444964-100 English Rev. 5