Table 3-3:
Product
LOGIQ E10
100-240 VAC
LOGIQ E9
100-240 VAC
LOGIQ P5
100-120 VAC
LOGIQ A5
200-240 VAC
LOGIQ E8
100-120 VAC
LOGIQ S8
200-240 VAC
LOGIQ S7
LOGIQ P3
100-120 VAC
200-240 VAC
LOGIQ P6
100-120 VAC
200-240 VAC
LOGIQ P9/P7
100-120 VAC
200-240 VAC
LOGIQ 9
100-120 VAC
200-240 VAC
100 VAC
LOGIQ 7
115 VAC
220 VAC
100 VAC
LOGIQ 5
115 VAC
230 VAC
100 VAC
LOGIQ 3
115 VAC
230 VAC
LOGIQ F Series
100-240 VAC
LOGIQ C9 Series
220-240 VAC
LOGIQ V3/LOGIQ
100-240 VAC
V5/LOGIQ V5
Expert
Ultrasound System
Direction 5444964-100 English Rev. 5
General Ultrasound system requirements
Electrical requirements - LOGIQ Consoles
Voltage
Tolerance
±10%
±10%
–
Common Service Information
Current or
Power
Consumption
0.9kVA
1100 W (valid
through R5)
1.0kVA (effective
from R6)
750 VA
900 VA
50/60Hz (±2Hz)
425 VA
950 VA
500 VA
9.5 A
4.75 A
MAX. 1200 VA
MAX. 1250 VA
MAX. 860VA
MAX. 400 VA
MAX. 950 VA
50/60Hz (±2Hz)
MAX. 300 VA
Frequency