GE LOGIC C2 Series Technical Publication page 117

Ultrasound system
Table of Contents

Advertisement

Generic procedure on probe leakage current (continued)
NOTE:
Table 4-16:
Typical data format for recording probe (source) leakage current
Unit under test___________________________
Test Conditions
System Power
off
closed
off
open
on
closed
on
open
NOTE:
Table 4-17:
Unit under test____________________________________
Test Conditions
System
Grounding/
Power
PE
off
closed
off
open
on
closed
on
open
off
closed
(reversed
polarity)
off
open (reversed
polarity)
on
closed
(reversed
polarity)
on
open (reversed
polarity)
Ultrasound System
Direction 5444964-100 English Rev. 5
Not all test procedures are applicable to all areas of the world.
Reversed Polarity testing content satisfies regions following IEC
62353:2007 and IEC 60601-1:2005.
Grounding /PE
Values in italics font are given as examples only.
Typical data format for recording probe (source) leakage current
Limit
µA
100
500
100
500
100
500
100
500
Common Service Information
Date of test:___________
Probe as measured in saline bath
4C
i12L
Probe as measured in saline bath
4C
i12L
Electrical safety tests
TS
E8C
Date of test:_____________
TS
E8C
4-39