Care and Maintenance
LOGIQ/Vivid/Voluson Ultrasound System Leakage Current Limits
Table 4-6:
Leakage current limits for Ultrasound system operation on 100-130 Volt
Leakage Current Test
Earth Leakage
Chassis/Enclosure Leakage
Type BF Applied Parts
Type CF Applied Parts
Type BF Applied Parts (sink
leakage, mains voltage on applied
part)
Type CF Applied Parts (sink
leakage)
(1) UL 60601-1 standard
NOTE:
4-20
Leakage current limits for LOGIQ, Vivid and VOLUSON
Ultrasound Systems are shown below for 100-130 Volt and
220-240 Volt mains.
mains (US/Canada/Japan)
System
Power
On and Off
On and Off
On (transmit)
On (transmit)
On and Off
On and Off
Open Grounding is also known as "Lift Ground".
Ultrasound System
Grounding/
PE Conductor
N/A
0.3
Closed
0.1
Open
0.3
Closed
0.1
Open
0.5
Closed
0.01
Open
0.05
Closed
5
Closed
0.05
Common Service Information
Direction 5444964-100 English Rev. 5
Limit in
Limit in
mA
µA
(1)
(1)
300
100
(1)
(1)
300
100
500
10
50
5000
50