Probe (Source) Leakage Current Test - GE LOGIC C2 Series Technical Publication

Ultrasound system
Table of Contents

Advertisement

Probe (Source) leakage current test

WARNING
Definition
NOTE:
Ultrasound System
Direction 5444964-100 English Rev. 5
Do not use the probe if the insulating material has been
punctured or otherwise compromised. Integrity of the insulation
material and patient safety can be verified by safety testing
according to IEC60601-1.
This test measures the current that would flow to ground from
any of the probes through a patient who is being scanned and
becomes grounded by touching some other grounded surface.
Each probe will have some amount of leakage, dependent on its
design. Small variations in probe leakage currents are normal
from probe to probe. Other variations will result from differences
in line voltage and test lead placement. It is abnormal if no
leakage current is measured. If no leakage current is detected,
check the configuration of the test equipment.
Common Service Information
Electrical safety tests
4-35