Analog Voltage Measurement Accuracy Offsets; Analog Voltage Measurement Speed - Campbell CDM-A100 Series Instruction Manual

Table of Contents

Advertisement

TABLE 6-1. Analog Voltage Measurement Accuracy Offsets
Differential with
Range
Input Reversal
(mV)
(µV)
±5000
±4
±1000
±2
±200
±1
TABLE 6-2. Analog Voltage Measurement Speed
With Input Reversal
f
Time
N1
(Hz)
(ms)
30000
1.46
15000
1.53
7500
1.66
3750
1.93
2000
2.40
1000
3.40
500
5.40
100
21.40
60
34.73
50
41.40
30
68.06
25
81.40
15
134.73
10
201.40
5
401.40
2.5
801.40
1
Default settling time of 500 µs.
2
Refers to multiplexing circuitry internal to the CDM-A100 series.
CDM-A100 Series
Differential or Single-
Ended without Input
Reversal
(µV)
±40
±10
±3
2
Multiplexed
Measurement
Without Input Reversal
Rate
Time
(Hz)
(ms)
698.49
0.75
667.41
0.78
612.87
0.85
526.78
0.98
422.83
1.22
297.18
1.72
186.39
2.72
46.81
10.72
28.82
17.38
24.18
20.72
14.70
34.05
12.29
40.72
7.42
67.38
4.97
100.72
2.49
200.72
1.25
400.72
1
Rate
(Hz)
1394.05
1332.15
1223.49
1051.89
844.59
593.82
372.58
93.60
57.63
48.35
29.40
24.58
14.85
9.93
4.98
2.50
13

Advertisement