Campbell CDM-A100 Series Instruction Manual page 59

Table of Contents

Advertisement

Appendix D. Analog Voltage Measurement Range and Resolution
D-2
Measurements with
Input Reversal
1
2
f
Range
N1
(Hz)
(mV)
RMS (µV)
±5000
60
±1000
±200
±5000
50
±1000
±200
±5000
30
±1000
±200
±5000
25
±1000
±200
±5000
15
±1000
±200
±5000
10
±1000
±200
±5000
5
±1000
±200
±5000
2.5
±1000
±200
1
First notch frequency
Range overhead of ≈6% on all ranges guarantees that full-scale values will not
2
cause overrange.
3
Effective resolution (ER) in bits is computed from ratio of full-scale range to RMS
noise.
Typical Effective Resolution
Single-Ended
Measurements or
Differential
Differential
Measurements without
Input Reversal
bits
RMS (µV)
0.769
23.7
1.140
0.162
23.6
0.261
0.056
22.9
0.113
0.732
23.8
1.112
0.161
23.7
0.254
0.053
22.9
0.111
0.612
24.1
0.901
0.131
24.0
0.217
0.042
23.3
0.099
0.580
24.1
0.858
0.123
24.0
0.204
0.038
23.4
0.095
0.502
24.4
0.734
0.107
24.2
0.177
0.031
23.7
0.088
0.499
24.4
0.683
0.108
24.2
0.161
0.029
23.8
0.082
0.472
24.4
0.617
0.102
24.3
0.151
0.024
24.1
0.082
0.447
24.5
0.564
0.095
24.4
0.144
0.020
24.3
0.077
3
bits
23.2
23.0
21.8
23.2
23.0
21.9
23.5
23.2
22.0
23.6
23.3
22.1
23.8
23.5
22.2
23.9
23.7
22.3
24.1
23.7
22.3
24.2
23.8
22.4

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cdm-a108Cdm-a116