Campbell CDM-A100 Series Instruction Manual page 52

Table of Contents

Advertisement

StdDev(1,Vr1000(2),IEEE4,False)
StdDev(1,Strain(3),IEEE4,False)
StdDev(1,Vr1000(3),IEEE4,False)
StdDev(1,Strain(4),IEEE4,False)
StdDev(1,Vr1000(4),IEEE4,False)
EndTable
DataTable(Table2,True,-1)
DataInterval(0,1440,Min,10)
Minimum(1,BattV,FP2,False,False)
EndTable
'Calibration history table
DataTable(CalHist,NewFieldCal,10)
SampleFieldCal
EndTable
'Main Program
BeginProg
'Initialize calibration variables for
'Quarter Bridge Strain, 3-wire 350 ohm with 4WFBS350 TIM measurement 'Vr1000()'
CIndex=1 : CAvg=1 : CReps=4
For
LCount = 1
To
4
GFAdj(LCount)=GFsRaw(LCount)
Next
'Load the most recent calibration values from the CalHist table
FCLoaded=LoadFieldCal(True)
'Main Scan
Scan(500,mSec,4,0)
'Default Datalogger Battery Voltage measurement 'BattV'
Battery(BattV)
'Default Wiring Panel Temperature measurement 'PTemp_C'
PanelTemp(PTemp_C,15000)
'Quarter Bridge Strain, 3-wire 350 ohm with 4WFBS350 TIM measurement 'Vr1000()'
CDM_BrFull(CDM_A116,10,Vr1000(),4,mV200,1,X1,3,2500,False,True,500,2000,1,0)
'Calculated strain result 'Strain()' for
'Quarter Bridge Strain, 3-wire 350 ohm with 4WFBS350 TIM measurement 'Vr1000()'
StrainCalc(Strain(),4,Vr1000(),BrZero(),-1,GFAdj(),0)
'Quarter bridge strain shunt calibration for
'Quarter Bridge Strain, 3-wire 350 ohm with 4WFBS350 TIM measurement 'Vr1000()'
FieldCalStrain(13,Strain(),1,GFAdj(),0,QBSSMode,CKnown(),CIndex,CAvg,GFsRaw(),0)
'Zeroing calibration for
'Quarter Bridge Strain, 3-wire 350 ohm with 4WFBS350 TIM measurement 'Vr1000()'
FieldCalStrain(10,Vr1000(),CReps,0,BrZero(),ZMode,0,CIndex,CAvg,0,Strain())
'Call Data Tables and Store Data
CallTable
Table1
CallTable
Table2
CallTable
CalHist
NextScan
EndProg
Appendix B. Example Programs
B-5