FIGURE 5-2. Power and CPI ports view

5.1.1 Analog Input

Analog sensors output a continuous voltage or current signal that varies with
the phenomena measured. Sensors compatible with the CDM-A100 output a
voltage. Current output can be made compatible with a resistive shunt.
Sensor connection is to H/L terminals configurable for differential (DIFF),
single-ended (SE), or period-average inputs. For example, differential channel
1 is comprised of terminals 1H and 1L, with 1H as high and 1L as low.

5.1.1.1 Single-Ended Measurements

A single-ended measurement measures the difference in voltage between the
terminal configured for single-ended input and the reference ground. Single-
ended terminals are labeled in blue. The CDM-A108 has 16 SE terminals, and
the CDM-A116 has 32. For example, single-ended channel 1 is comprised of
terminals 1 and ⏚.
Maximum input voltage range: ±5000 mV
Measurement resolution range: 0.020 to 14.756 µV
CDM-A100 Series
Power In
CPI Ports
CPI Port LED
Indicators
RMS