7.5 Using shaped composite seals

Fig. 23: Moulded composite seals with
grooved insert
Fig. 24: Elastomer flat seal with a smooth
insert

7.6 Replacing the safety relief valve

Leaks are possible
Leaks can occur on the pump connection depending on
the insert used.
The pump is supplied with PTFE moulded compo‐
site seals with a flare, which are used for the pump
connectors, and which seal the connectors between
the grooved pump valves and the grooved inserts -
P_SI_0021
see .
However, use an elastomer flat seal in the event that
a smooth insert is used (e.g. third party part), .
P_SI_0022
WARNING!
The safety relief valve is a safety-relevant component
and should only be fully replaced if necessary.
1.
Remove the old safety relief valve.
2.
Pay attention to the position of the new O-ring. Fix the O-ring in
place with a little grease if necessary.
3.
Screw in the new safety relief valve (tightening torque - 15 Nm).
4.
Mark the collar of the safety relief valve with sealing paint.
Maintenance
39