Safety Inspections; Demounting; Accessories - Derungs Dmed Triango 100 Installation And Operating Instructions Manual

Treatment lights
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 23
ENG
10

SAFETY INSPECTIONS

DANGER
Electric shocks are life-threatening.
Deenergize the device/pull out the mains plug
and secure against being switched on again.
CAUTION
Maintenance and repairs can only be per-
formed by qualified electricians.
The corresponding user profile is in Section 2
Safety instructions.
annually:
• Check the power supply cable for damage and
replace if necessary
• Check the fastening screw on the underside
tripod foot
and tighten if necessary.
• Paint damage
• Cracks on plastic parts
• Deformation of the carrier system
• Loosening of parts
10.1
Setting the spring force
CAUTION
The spring force is set at the optimum setting in
the factory.
4
metric
As a first measure slightly tighten the joint
cover (by a ¼ turn); this to achieve greater
stability of the luminaire head, if necessary
Otherwise:
Remove the service cover of the spring
balanced arm (loosen the 2 crosstip screws)
Place a suitable tool (e.g. a 3-type Allen key)
into the hole 1 , turn by ¼ turn in the direction
of the arrow, a repeat a number of times until
the desired stability is reached
34
1
Screw the service cover on again (attention:
do not tighten too strongly)
11

DEMOUNTING

DANGER
Electric shocks are life-threatening.
Disconnect the power connection from the
network
DANGER
Risk of injury
The spring balanced arm is under a high
spring force. If the terminal device is not in
the uppermost spring balanced arm posi-
tion, the spring balanced arm will move rap-
idly upwards and lead to severe injuries.
Only disassemble the terminal device when
the spring balanced arm is in the uppermost
position
11.1
Disposal
Do not place the luminaire in the
household waste. Bring the luminaire
and lamp, according to local regula-
tions, to a disposal site or give it to a
dealer with the appropriate service
offering.
Cut the cable directly at the casing.
The products listed above are over
95% recyclable. In order for a high
percentage of the used materials to
either be physically re-used or used for
energy after the end of this product's
life cycle, the luminaires have been
designed with recycling in mind. They
do not contain hazardous or supervisi-
on-requiring substances.
12

ACCESSORIES

Ergonomic handle (order no. D10.295.000)
Handle cover (order no. D15.445.000)

Hide quick links: