LUMINAIRE HEAD ASSEMBLY

WARNING
Risk of injury from a falling luminaire head
Ensure that the luminaire head is secured in
an orderly fashion.
Damage to due to a poorly mounted luminaire
head.
Connect the luminaire head plug to the wall-
mounted joint plug
Before assembly one must ensure that the
brass nose (see arrow) and the cardan joint
bracket are on the left
2.5
metric
2
.
2
Push together the luminaire head 1 and arm 3;
in doing so the cardan joint bracket must be on
the left 2 of the luminaire head
then screw in both countersunk screws M4
(both with a toothed washer) in (1. 5Nm)
Check the turning function
Plug in the handle
32
m
e
t
8

OPERATION

WARNING
Warning against damage to eyes
Never look directly into the light cone.
Operating function (dependent on the model)
1
2
3
1
2
Switch the luminaire on or off using the button
1.
The luminaire can be dimmed using the +/-
buttons 2
The light color can be adjusted at 3
The focus can be adjusted at 4
The mode of use of the endoscope can be
adjusted at 5
Before each use, perform a function test: all
LEDs in the light cone must come on.
Triango 100 F
DANGER
Electric shocks are life-threatening.
Do not plug in any damaged power cables.
If there are any signs of damage to the
power cable, immediately replace it with a
new one.
Connection voltage and frequency must
match data on the type plate.
Only connect the luminaire to the supply
network with a protective earth conductor
(PE).
CAUTION
Triango 100 F
Wind the power cable onto the cable holder
when not in operation
Plug in cable
Connect cable to the network
4
5