CLEANING

DANGER
Danger to life through electric shock
Before cleaning: disconnect the power
connection from the network
CAUTION
Material damage due to incorrect cleaning
For cleaning, only use agents which do not
affect functioning of the luminaire.
For cleaning, do not use any solvent or
chlorine based or abrasive detergents as
they can, among other things, result in
cracking of plastic parts.
The cleaning agents used must be ap-
proved for use on plastics such as PC,
PMMA, PA and ABS.
Damage to the luminaire due to concen-
trated disinfectant.
For concentration and application times,
please consult the information provided
with the agent used.
Use of the wrong cloths can make scratch-
es
CAUTION
Dirt reduces the luminosity
Keep cover clear through regular cleaning.
Only wipe cleaning allowed.
Clean the PA clear cover using a non-
abrasive cleaning cloth and a suitable clean-
ing agent
Wipe cleaning is only allowed in a horizontal
position
Recommended cleaning agent
Use a mild soap solution or commercially available
detergent as a cleaning agent. Wipe the surfaces
of the devices with a slightly moistened cloth; add a
mild soap solution (detergent) if necessary. Finally
wipe dry the outer surfaces using a soft, clean
cloth.
CAUTION
Clean external luminaire parts using the follow-
ing prescribed diluted products:
Lysoformin
®
Dismozon
®
Hexaquart
plus
®
Sagrotan
quick disinfectant cleaner
®
CAUTION
To minimize the risk of disease transmission, in
addition to complying with this user manual, you
must also comply with the applicable occupa-
tional health and safety regulations and the
requirements of national bodies with responsibili-
ties for hygiene and disinfection.
9.1
Sterilization of the ergonomic
handle
Sterilization must be carried out in accordance
with ISO 17665-1 (Sterilization of health care
products in moist heat).
CAUTION
Damage to the hand grip
Do not sterilize with hot air
Package the hand grip in a sterile bag before
sterilization.
The hand grip is designed exclusively for
damp sterilization with 3 times fractioned pre-
vacuum and saturated steam with the follow-
ing parameters:
Temperature
Overpressure
Dwell time
Vacuum drying
WARNING
Warning against burns
The ergonomic handle must be cooled to
room temperature before use.
After sterilization check the ergonomic handle
for mechanical integrity; do not continue to use
damaged
ergonomic
ENG
134°C
2.0 bar
6 min
20 min
handles
33