2 E
QUIPMENT
P
I
YCNOMETER
NSTALLATION
The AccuPyc is designed to operate with a power supply of 90-264V~ at 50/60Hz. Noise-free power
of the correct voltage and frequency, with a safety earth ground, should be available through a
standard wall receptacle. The power outlet should be able to supply 15 amps @ 100 or 115V~ ±10%
or 7.5 amps @ 230V~ ±10%. There should also be sufficient outlets for all devices.
The external power adapter required for the AccuPyc is Micromeritics' part number 003-
40054-00. Use of any other power adapter could damage equipment and/or cause harm
to the operator. The AccuPyc is intended to be powered from the output of the approved
power adapter rated Class III, manufacture by TRUMPower, P/N TSA42-D21 (R1).
Micromeritics supplies a suitably rated approved power supply cord appropriate for the
applicable country with the power adapter.
AccuPyc II 1345 Installation Instructions and Checklist
134-42875-02 (Rev A) — Jan 2021
I
NSTALLATION

2 Equipment Installation

2 - 1