JDS Uniphase cORL-A1 Operating Manual page 52

Optical return loss meter
Table of Contents

Advertisement

7 R
C
EMOTE
Command
string
:POW
:CAL
:WAV
:TAB?
:POW
:CAL
:WAV
:POW
:CAL
:WAV?
:POW
:DISP
:UNIT
:POW
:DISP
:UNIT?
:POW
:REF
:STAT
48
ONTROL
Parameter type / Response type /
Unit / Info
Returns the contents of the λ-Table.
Response type: <NR1,NR1,...>
Selects the calibration wavelength
from the λ-Table.
See :SYST:CAL:WAV:TAB? for more
details.
Parameter type: <NR1>
e.g. :POW:CAL:WAV 850
Returns the calibration wavelength
currently used by the instrument.
e.g. 1310
Selects how the power level is
displayed.
Parameter type: <MNEMONIC>
• LIN: linear display of power level
(µW)
• LOG: logarithmic display of power
level (dBm or dB)
Returns the power level display type:
LIN or LOG
Determines whether the power level
reading is displayed as an absolute or
a relative value.
Parameter type: <BOOLEAN>
• 0: absolute value (µW or dBm)
• 1: relative value (dB)
Default setting: 0
cORL-A1
JDSU