7 R
C
EMOTE
Command
string
:SOUR
:WAV
:AUTO
:SOUR
:WAV
:AUTO?
:SOUR
:MOD
:SOUR
:MOD?
:SOUR
:POW
46
ONTROL
Parameter type / Response type /
Unit / Info
Sets Auto-λ modulation ON/OFF.
Parameter type: <BOOLEAN>
• 0: Auto-λ OFF
• 1: Auto-λ ON
Default setting: OFF (0)
Returns Auto-λ modulation status.
• 0: Auto-λ OFF
• 1: Auto-λ ON
Sets the modulation for the selected
LASER source.
Parameter type: <NR1>
Values allowed:
• 0: CW (continuous wave)
• 270: mod. freq. 270 Hz
• 1000: mod. freq. 1000 Hz
• 2000: mod. freq. 2000 Hz
Note: Setting will fail, if more than one
LASER source is selected.
Returns the current modulation.
• 0: CW (continuous wave)
• 270: mod. freq. 270 Hz
• 1000: mod. freq. 1000 Hz
• 2000: mod. freq. 2000 Hz
• AUTO: Auto-λ modulation is ON.
Sets the power level for the specified
LASER source.
Parameter type: <NR1,NR1>
e.g. :SOUR:POW 1,-522
sets the power level for LASER1 to -
5.22 dBm