Maintaining the debubbler and measuring
chamber
8.3.1
Clean the debubbler and measuring chamber
Procedure
1. Turn off the sample supply to the debubbler.
2. Remove the sensor and put it in a safe place.
3. Loosen the small drain plug in the base plug and allow the sample in the debubbler
4. Replace the drain plug.
5. Unscrew the upper and lower caps.
6. Use a stream of water, a brush, or a rag to flush and clean out the inside of the
7. Inspect the O-rings for signs of damage and replace if necessary.
8. Replace the upper and lower caps.
9. Replace the sensor.
8.3.2
Clean the debubbler orifice
Procedure
1. Turn off the sample to the debubbler.
2. Disconnect the drain line. Unscrew the drain fitting from the orifice; then unscrew
3. Use a stream of water to flush out any residue accumulated in the orifice. Direct the
Rosemount T56
WARNING
High pressure and temperature
Before disconnecting the sample and drain lines or removing the sensor, be
absolutely certain the process pressure is reduced to 0 psig and the process
temperature is at a safe level.
The calibration cup is a good place to store the sensor.
to drain out.
See
Figure
8-1.
Be careful not to lose the O-rings.
debubbler and measuring chamber.
The part number for the O-ring (one each) is 9550316.
the orifice from the debubbler body.
See
Figure
8-1.
stream of water counter to the normal flow through the orifice.
Maintenance
July 2020
53