Download Print this page
Emerson Rosemount Clarity II T56 Quick Start Manual
Emerson Rosemount Clarity II T56 Quick Start Manual

Emerson Rosemount Clarity II T56 Quick Start Manual

Turbidmeter
Hide thumbs Also See for Rosemount Clarity II T56:

Advertisement

Quick Links

Quick Start Guide
00825-0100-3565, Rev AA
April 2020
Rosemount Clarity II T56 Turbidmeter

Advertisement

loading

Summary of Contents for Emerson Rosemount Clarity II T56

  • Page 1 Quick Start Guide 00825-0100-3565, Rev AA April 2020 Rosemount Clarity II T56 Turbidmeter...
  • Page 2 Essential instructions Read this page before proceeding! Your instrument purchase from Emerson is one of the finest available for your particular application. These instruments have been designed and tested to meet many national and international standards. Experience indicates that its performance is directly related to the quailty of the installation and knowledge of the user in operating and maintaining the instrument.
  • Page 3 Warranty 1. LIMITED WARRANTY: Subject to the limitations contained in Section 2 herein and except as otherwise expressly provided herein, Emerson (“Seller”) warrants that the firmware will execute the programming instructions provided by Seller and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expiration of the applicable warranty period.
  • Page 4 INCLUDE INCIDENTAL, CONSEQUENTIAL, OR PUNITIVE DAMAGES. THE TERM “CONSEQUENTIAL DAMAGES” SHALL INCLUDE, BUT NOT BE LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE, AND COST OF CAPITAL. Contents Install............................5 Wire............................22 EU Declaration of Conformity.....................27 China RoHS table........................29 Emerson.com/Rosemount...
  • Page 5: Install

    April 2020 Quick Start Guide Install Unpack and inspect Consult the table to verify that you have received the parts for the option you ordered. Table 1-1: Rosemount Clarity II Turbidmeter Parts Item Model/part number Single input turbidity transmitter with 56-03-27-38-HT ®...
  • Page 6 April 2020 WARNING Electrical shock Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other applicable national or local codes. Do not operate or energize instrument with case open. Figure 1-1: Panel Mount Front View Emerson.com/Rosemount...
  • Page 7 April 2020 Quick Start Guide Figure 1-2: Panel Mount Side View A. Panel mount gasket B. Panel supplied by others. Maximum thickness: 0.375 in. (9.52 mm) C. 4X mounting brackets and screws provided with instruments Quick Start Guide...
  • Page 8 Quick Start Guide April 2020 Figure 1-3: Panel Mount Bottom View A. Conduit openings Emerson.com/Rosemount...
  • Page 9 April 2020 Quick Start Guide Figure 1-4: Panel Cut-Out A. Maximum Quick Start Guide...
  • Page 10 Quick Start Guide April 2020 Figure 1-5: Wall/Surface Mount Front View A. 2-in. pipe supplied by customer. Vertical pipe mounting shown. For horizontal pipe, rotate brackets and U-bolts 90 °. Note Unless otherwise specified. B. 4X cover screw Emerson.com/Rosemount...
  • Page 11 April 2020 Quick Start Guide Figure 1-6: Wall/Surface Mount Side View Quick Start Guide...
  • Page 12 Quick Start Guide April 2020 Figure 1-7: Pipe Mount Bottom View A. Front panel B. 2-in. pipe mount bracket C. 2X set U-bolts for 2-in. pipe in kit PN 23820 Emerson.com/Rosemount...
  • Page 13 April 2020 Quick Start Guide Figure 1-8: Pipe Mount Side View Quick Start Guide...
  • Page 14 2. Attach a piece of ⅜-in. ID soft tubing to the drain fitting. The debubbler must drain to atmosphere. WARNING High pressure and temperature Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level! Emerson.com/Rosemount...
  • Page 15 April 2020 Quick Start Guide CAUTION Reading errors During operation, the debubbler is under pressure. A 0.040 in. (1 mm) orifice in the outlet provides the pressure. Back pressure helps prevent outgassing, which can lead to bubbles accumulating on the sensor face, resulting in erroneous readings.
  • Page 16 The sensor is rated to IP65 when properly connected to the cable. Postrequisites To prevent possible water damage to the connector contacts, be sure the cable receptacle and the connector on the back of the sensor are dry when connecting or disconnecting the cable. Emerson.com/Rosemount...
  • Page 17 April 2020 Quick Start Guide Install Rosemount 56 transmitter Figure 1-11: Non-Incendive Field Wiring Installation for the Rosemount 56-27-37 Transmitter A. Metal conduit B. If sensor is equipped with a VP connector, the VP connector must be in the unclassified area. C.
  • Page 18 Class I, Division 2, Groups A-D, 0 ° to 50 °C Class II, III, Division 2, Groups E-G K. Alarm wiring (Vac) (optional) L. Analog output (optional) M. Ground connection may be made in a hazardous area. N. Power supply Emerson.com/Rosemount...
  • Page 19 April 2020 Quick Start Guide Figure 1-12: CSA Non-Incendive Field Wiring Installation A. Metal conduit B. Recommended sensor attached. See Drawing 1700704 for sensors recommended for use with option -31. (No sensor if blank or -38). WARNING Use with non-flammable process media only. D.
  • Page 20 K. Alarm wiring (Vac) (optional) L. Analog output (optional) M. Power supply N. You may make a ground connection in a hazardous area. Note A. Installation must conform to the CEC. B. Seal required at each conduit entrance. C. Unless otherwise specified. Emerson.com/Rosemount...
  • Page 21 April 2020 Quick Start Guide Figure 1-13: Non-Incendive Field Wiring Connection for Class 1, Division 1, Group D Turbidity sensor board Option -27/-37: turbidity May only be used with a Rosemount Clarity II Turbidity Sensor. Quick Start Guide...
  • Page 22: Wire

    HART communication supports Bell 202 digital communications over an analog 4-20 mA current output. Profibus DP is an open communications protocol which operates over a dedicated digital line to the host. Emerson.com/Rosemount...
  • Page 23 2.1.4 Alarm relays Emerson supplies four alarm relays with the switching power supply (85 to 264 Vac, 03 order code) and the 24 Vdc power supply (20 - 30 Vdc, 02 order code). You can use all relays for process measurement(s) or temperature.
  • Page 24 2.4.3 Alarm relay wiring Emerson supplies four alarm relays with the switching power supply (85 to 265 Vac, 03 order code) and the 24 Vdc power supply (20-30 Vdc, 02 order code). Wire the relay leads on each of the independent relays to the correct position on the power supply board using the printed lead markings (NO/ Normally open, NC/Normally closed, or Com/Common) on the board.
  • Page 25 April 2020 Quick Start Guide Figure 2-1: Turbidity Signal Board with Plug-in Sensor Connection The cable also passes through a strain relief fitting. To install the cable: Procedure 1. Remove the wrenching nut from the strain relief fitting. 2. Insert the plug through the hole in the bottom of the enclosure nearest the sensor socket.
  • Page 26 Cycle power to the instrument after connecting to the sensor or • Perform a slope calibration or standard calibration routine after connecting the sensor. Following these guidelines will extend the life of the incandescent lamp and avoid premature warnings and faults due to reduced lamp life. Emerson.com/Rosemount...
  • Page 27: Eu Declaration Of Conformity

    April 2020 Quick Start Guide EU Declaration of Conformity Quick Start Guide...
  • Page 28 Quick Start Guide April 2020 Emerson.com/Rosemount...
  • Page 29: China Rohs Table

    April 2020 Quick Start Guide China RoHS table Quick Start Guide...
  • Page 30 Quick Start Guide April 2020 Emerson.com/Rosemount...
  • Page 31 April 2020 Quick Start Guide Quick Start Guide...
  • Page 32 LinkedIn.com/company/Emerson- Automation-Solutions The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a Twitter.com/rosemount_news mark of one of the Emerson family of companies. Facebook.com/Rosemount All other marks are the property of their Youtube.com/RosemountMeasurement respective owners.