Important
When using EPA/incandescent sensors (P 8-0108-0002-EPA):
Do not power up the instrument without the sensor connected.
Do not disconnect and reconnect a sensor while a transmitter is powered.
If this is inconvenient or cannot be avoided:
Cycle power to the instrument after connecting to the sensor or
Perform a slope calibration or standard calibration routine after connecting the sensor.
Following these guidelines will extend the life of the incandescent lamp and avoid
premature warnings and faults due to reduced lamp life.
Rosemount T56
Note
If S1 Warning appears, check sensor cable connection and confirm sample water
flow at debubbler drain outlet.