Emerson Rosemount Clarity II T56 Reference Manual page 47

Turbidmeter
Hide thumbs Also See for Rosemount Clarity II T56:
Table of Contents

Advertisement

Reference Manual
00809-0100-3565
7.2.3
Calibrate against a grab sample
If you want, you can calibrate the turbidity sensor against the turbidity reading from
another instrument.
The transmitter treats the value you entered as though it were the true turbidity of the
sample. Therefore, grab sample calibration changes the sensitivity; it does not apply an
offset to the reading.
Rosemount T56
Calibrate
July 2020
47