Figure 3-1: Turbidity Signal Board with Plug-in Sensor Connection
The cable also passes through a strain relief fitting. To install the cable:
Procedure
1. Remove the wrenching nut from the strain relief fitting.
2. Insert the plug through the hole in the bottom of the enclosure nearest the sensor
3. Slide the wrenching nut over the cable plug and screw it onto the fitting.
4. Loosen the cable nut so the cable slides easily.
5. Insert the plug into the appropriate receptacle. To remove the plug, squeeze the
6. Adjust the cable slack in the enclosure and tighten the cable nut.
7. Plug the cable into the back of the sensor.
8. Place the sensor in either the measuring chamber or the calibration cup.
28
socket. Seat the fitting in the hole.
release clip and pull straight out.
Be sure to allow sufficient slack to avoid placing stress on the cable and
connections.
The sensor is rated to IP65 when properly connected to the cable. To prevent
possible water damage to the connector contacts, be sure the cable receptacle and
the connector on the back of the sensor are dry when connecting or disconnecting
the cable.
Important
The sensor must be in a dark place when power is first applied to transmitter.