Figure 5-3: Reset Screen
To reset settings, press ENTER/MENU from the main screen to access the Reset screen.
5.6
Program alarm relays
The Rosemount
order code) provide four alarm relays for process measurement or temperature. You can
program each relay independently as an interval timer or one of four advanced timer
functions.
You can program the following relay functions to any relay from the Relays screen.
•
Assign a relay.
•
Define a relay function.
•
Assign a measurement.
•
Set relay logic.
•
Enter setpoints.
•
Set deadband.
•
Set normal state.
•
Set USP safety level (contacting conductivity).
Rosemount T56
™
T56 24 Vdc (-02 order code) and the AC switching power supply (-03
Program the transmitter