Chapter 6 Program Turbidity - Emerson Rosemount Clarity II T56 Reference Manual

Turbidmeter
Hide thumbs Also See for Rosemount Clarity II T56:
Table of Contents

Advertisement

Reference Manual
00809-0100-3565
6
Program turbidity
The Rosemount
time you power up the transmitter. Upon first power-up, completing Quick Start screens
enables measurements, but you may need to complete additional steps to program the
transmitter for the desired measurement application.
The following programming and configuration functions are covered:
Measurement
Sensor type
Units
Bubble rejection
Filter
Figure 6-1: Measure Tab
Procedure
1. To program the Measurement type, select Turbidity or TSS calculation (estimated
2. To program the Sensor type, select EPA or ISO and press ENTER/MENU.
3. To program Units, select NTU, FTU, or FNU and press ENTER/MENU.
4. To program Bubble rejection, select On or Off and press ENTER/MENU.
5. To override the default input Filter, enter 0 - 999 seconds and press ENTER/MENU.
Rosemount T56
T56 automatically recognizes each installed measurement board each
TSS) and press ENTER/MENU.
Program turbidity
July 2020
43